ADAM CellT Plus QC Slide

QC Slide for ADAM™ CellT Plus

The QC slide is used for “QC mode” of ADAM™ CellT Plus to monitor the accuracy and viability of the instrument.
Additionally, it reduces concerns of customers by checking the accuracy of the instrument frequently. 


QC Slide is used

- To provide absolute particle numbers

- To calibrate the automatic focus

- To check the position of the slide stage


 




 
 
 


 

Quality Control Protocol 


Performance test with QC Slide 


1) Activate the ADAM™ CellT Plus QC mode using QC slide lot number and activation code.

2) Please register the QC Slide lot in the ‘QC’ tab QC slide Edit.

3) Insert the QC Slide and operate the instrument.

4) Check the result on the ADAM™ CellT Plus QC mode screen or test report. 



How to correctly insert QC Slide


Make sure to insert the QC Slide correctly following the direction of arrow.










Data Analysis



 


  

1) The result of performance test2 should meet the ‘Acceptance range' in the QC slide.

(Example: 1.80E5 ~ 2.20E6 /mL)

2) Acceptance peak size check. (ADAM™ CellT Plus peak graph value: 13 ~16 μm).

a. Click the ‘QC’ tab-‘SAVE’ or ‘MAIL’ button on ADAM-CellT Plus screen.

b. Check the QC Slide peak size (See Cell Size Information) in the exported ADAM™ CellT Plus test report.









FOR RESEARCH USE ONLY.

This product is not approved for diagnostic or therapeutic use.




  • Product name / Cat. No.

    QCS–002

  • Acceptance range

    1.8 × 10E6–2.2 × 10E6/mL

  • Acceptance slide peak size

    13–16 μm

  • Dimensions

    74.9 mm (W) x 24.9 mm (D) x 2.0 mm (H)

  • Expiration date

    1 year