Home / Product / Detail page

QC plate for EVE™ HT


QC plate is used for “QC mode” of EVE™ HT to monitor the accuracy and viability of the instrument.
It is used to validate functionality of the EVE™ HT with repeatability, so it doesn’t require further adjustment by the customers.
QC Plate is used,

Additionally, it reduces concerns of customers by checking the accuracy of the instrument frequently.
Select the ‘Data’ tap to check the results. The result of performance test should meet the ‘Acceptance range' in the QC Plate. - Acceptance range: ㆍLow level (Cat. no. EHPQ-001) Cell concentration : 6.96 x 104 ~ 1.04 x 105 cells/mL Viability : 25.0 ~ 35.0 % ㆍMiddle level (Cat. no. EHPQ-002) Cell concentration : 9.00 x 105 ~ 1.10 x 106 cells/mL Viability : 45.0 ~ 55.0 % ㆍHigh level (Cat. no. EHPQ-003) Cell concentration : 4.50 x 106 ~ 5.50 x 106 cells/mL Viability : 75.0 ~ 85.0 % |
|
Want to learn more about cell counting and cell therapy?
Visit to find out more!
Product name / Cat no.
EHPQ–001 (Low level), EHPQ–002 (Middle level), EHPQ–003 (High level)
Dimensions
134 mm (W) x 92 mm (D) x 2.1 mm (H)
Expiration date
3 years from date of manufacture
ExTransfection™ provides open protocols that make it easy for users to set up and run experiments. Its 24-well optimization method helps users to optimize transfection protocols easily and quickly.
ExTransfection™ also offers a pre-programmed optimization protocol that helps an easy and quick optimization of electrical parameters for both adherent and suspended cells.